80 - 190 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-750

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-26 for 80 - 190 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-750 manufactured by Charter Medical, Ltd..

Event Text Entries

[112455798] A single, used bag was received for investigation. The bag was visually examined and no issues were observed. The bag was leak tested and a film pinhole was verified near the perimeter seal at the top portion of the bag. Visual examination of the bag under a magnification inspection lamp verified a small hole in the bag film. No other issues were observed on the bag. There is no information to suggest the hole existed at the time the bag was produced at the manufacturing facility. Other than the hole itself, there was no evidence of excessive force or damage to the bag, or associated ports and tubing. It is unknown if a manufacturing issue, or user facility issue, may have contributed to the pinhole, however there was no direct evidence of a root cause.
Patient Sequence No: 1, Text Type: N, H10


[112455799] The cf-750 was removed from its packaging pouch and filled with 100ml of a cell product. As the bag was being packed into a cassette, a leak was detected. There was no patient contact or infusion of cell product from the bag associated with this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00013
MDR Report Key7637288
Date Received2018-06-26
Date of Report2018-06-26
Date of Event2018-05-30
Date Mfgr Received2018-05-30
Device Manufacturer Date2018-01-15
Date Added to Maude2018-06-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name80 - 190 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name80 - 190 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2018-06-26
Returned To Mfg2018-06-08
Model NumberCF-750
Catalog NumberCF-750
Lot Number151751
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-26