VITEK? 2 GRAM NEGATIVE ID TEST KIT 21341

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-06-26 for VITEK? 2 GRAM NEGATIVE ID TEST KIT 21341 manufactured by Biomerieux Inc..

Event Text Entries

[112847064] A customer from the united states notified biom? Rieux of misidentification results when testing a patient isolate, obtained from a urine sample, with the vitek? 2 gn id test kit (ref 21341), lot 2410505203. The isolate was tested three times with the same lot of vitek 2 gn cards. Each time, vitek 2 provided an identification result of klebsiella oxytoca. The isolate was tested three times with vitek? Ms, and each time an identification of klebsiella pneumonia was obtained. The customer also performed an indole test, which was negative, indicating the strain is klebsiella pneumoniae. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2018-00237
MDR Report Key7638121
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-06-26
Date of Report2018-08-29
Date Mfgr Received2018-08-04
Device Manufacturer Date2018-04-15
Date Added to Maude2018-06-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. CANDACE MARTIN
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX INC.
Manufacturer Street595 ANGLUM ROAD
Manufacturer CitySAINT LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 GRAM NEGATIVE ID TEST KIT
Generic NameVITEK? 2 GN ID TEST KIT
Product CodeLQM
Date Received2018-06-26
Catalog Number21341
Lot Number2410505203
Device Expiration Date2019-04-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX INC.
Manufacturer Address595 ANGLUM ROAD SAINT LOUIS MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-26

© 2026 FDA.report