CODMAN SKULL TONG 66-1025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2006-09-22 for CODMAN SKULL TONG 66-1025 manufactured by Codman & Shurtleff, Inc..

Event Text Entries

[499806] Rep reports that the product fractured the patients skull.
Patient Sequence No: 1, Text Type: D, B5


[7834249] Upon completion of the investigation it was noted that this complaint was non-verifiable. Evaluation of the three skull tongs received indicated that one of the tongs was assembled correctly. This tong has a spring-loaded point (screw) and a solid point (screw) in place. The other two tongs have two solid point screws (1 tong) and 2 spring loaded screws (1 tong). It appears that the instruments have been re-assembled incorrectly. Incorrect assembly may contribute to poor instrument function. These instruments are inspected during the manufacturing process, incorrect assembly would have been detected during inspection. There were no metallurgic defects such as corrosion on the metal surfaces. All screws demonstrated well formed threads and were fully functional. This complaint is associated with improper assembly in the clinical setting and is not attributed to manufacturing processes. Corrective action is not deemed necessary based on the analysis results. Trends will be monitored for this and/or smaller complaints. At the present time this complaint is considered closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226348-2006-00254
MDR Report Key763944
Report Source07
Date Received2006-09-22
Date Mfgr Received2006-08-25
Date Added to Maude2006-09-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMATTHEW KING
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088283106
Manufacturer G1CODMAN & SHURTLEFF, INC
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal Code02767
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCODMAN SKULL TONG
Generic NameVARIOUS SKULL TONGS FOR TRACTION
Product CodeHAX
Date Received2006-09-22
Returned To Mfg2006-09-05
Model NumberNA
Catalog Number66-1025
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key751820
ManufacturerCODMAN & SHURTLEFF, INC.
Manufacturer Address325 PARAMOUNT DRIVE RAYNHAM MA 02767 US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2006-09-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.