30G HYDRODISSECTION CAN.MOD. G-34270

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-27 for 30G HYDRODISSECTION CAN.MOD. G-34270 manufactured by Aspen Surgical Products, Caledonia.

Event Text Entries

[112563775] Aspen surgical received a report from the distributor indicating that a 30g cannula was found with an incomplete seal on the blister container. A sample was provided for evaluation along with manufacturing lot number. A review of the sample confirmed the issue as indicated from the distributor. Analysis of the finished good lot number was reviewed. No non-conformance's were noted during the manufacturing process. The blister container appeared to be fully formed and showed no signs of damage. The seal weld at the peel open flap however, revealed signs of an incomplete seal. Based on the shape of the seal weld, the inner pouch was incidentally sealed in the outer pouch causing it to be incomplete. Therefore, the reported issue was the result of a loading error and a root cause for the incomplete seal is attributed to operator error. No further action is required as aspen surgical no longer manufactures or packages this part number.
Patient Sequence No: 1, Text Type: N, H10


[112563776] Aspen surgical received a report from the distributor indicating that a 30g cannula was found with an incomplete seal on the blister container. The item was not in use and no injury/death was reported. This report was filed in our complaint handling system as (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1836161-2018-00077
MDR Report Key7642470
Date Received2018-06-27
Date of Report2018-05-28
Date of Event2018-05-28
Date Mfgr Received2018-05-28
Device Manufacturer Date2017-06-12
Date Added to Maude2018-06-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JORDAN HACKERT
Manufacturer Street6945 SOUTHBELT DR. S.E.
Manufacturer CityCALEDONIA MI 49316
Manufacturer CountryUS
Manufacturer Postal49316
Manufacturer Phone6165367508
Manufacturer G1SAME AS ABOVE
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name30G HYDRODISSECTION CAN.MOD.
Generic NameOPHTHALMIC CANNULA
Product CodeHMX
Date Received2018-06-27
Model NumberG-34270
Lot Number138466
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerASPEN SURGICAL PRODUCTS, CALEDONIA
Manufacturer Address6945 SOUTHBELT DR. S.E. CALEDONIA MI 49316 US 49316


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-27

© 2026 FDA.report