GLUMA DESENSITIZER 65872354

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-06-28 for GLUMA DESENSITIZER 65872354 manufactured by Kulzer Gmbh.

Event Text Entries

[112698404] As allowed by exemption# (b)(4), (b)(4) is submitting the report on behalf of kulzer (b)(4) (the manufacturer). This is a serious injury (as defined in 21 cfr section 803. 3) this incident is being reported to maintain compliance with 21 cfr 803 and out of an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


[112698405] Dr called and he has a patient in the chair that had gluma applied to a class 5 restoration at the cervical root surface 2 days ago. The application was for 3 minutes and then the material was rinsed off. It is unclear as to whether isolation was used. The patient has suffered a tissue burn and the tissue is black in color and they are in pain. Tissue burns are known side affect of gluma if the material contacts soft tissues. The dfu recommends the use of a rubber dam for isolation to protect the soft tissues. Dfu also states an application time of 30-60 seconds, using the smallest amount of material as possible.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610902-2018-00004
MDR Report Key7648409
Report SourceHEALTH PROFESSIONAL
Date Received2018-06-28
Date of Report2018-06-28
Date of Event2018-05-30
Date Facility Aware2018-06-01
Date Mfgr Received2018-06-01
Date Added to Maude2018-06-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS AMBER BROWN
Manufacturer Street4315 S. LAFAYETTE BLVD
Manufacturer CitySOUTH BEND IN 46614
Manufacturer CountryUS
Manufacturer Postal46614
Manufacturer Phone5472995411
Manufacturer G1KULZER, GMBH
Manufacturer StreetPHILIPP-REIS-STRASSE 8/13
Manufacturer CityWEHRHEIM, D-61273
Manufacturer CountryGM
Manufacturer Postal CodeD-61273
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGLUMA DESENSITIZER
Generic NameDESENSITIZER
Product CodeKLE
Date Received2018-06-28
Catalog Number65872354
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerKULZER GMBH
Manufacturer AddressPHILIPP-REIS-STRASSE 8/13 WEHRHEIM, D-61273 GM D-61273


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-28

© 2026 FDA.report