INTROCAN? SAFETY N/A 4251128-03

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-06-29 for INTROCAN? SAFETY N/A 4251128-03 manufactured by B. Braun Melsungen Ag.

Event Text Entries

[112718285] (b)(4). Device history record review (dhr): reviewed the device history record and no abnormalities found during in process and final control inspection. Preliminary investigation: received one used cannula hub (contaminated with blood) of introcan safety 3-w pur 22g 0. 9x25 mm-ap without packaging. The capillary hub and protective cap were not returned for investigation. Checked the returned sample visually and observed that the safety clip was located towards the tip of the needle (not in engaged position). Additionally, the needle was also inspected and no defect or damage was detected. Forwarded the sample to production for further investigation and evaluation.
Patient Sequence No: 1, Text Type: N, H10


[112718286] As reported by the user facility ((b)(4)): safety clip came off.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610825-2018-00118
MDR Report Key7649593
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-06-29
Date of Report2018-11-27
Date of Event2018-05-16
Date Facility Aware2018-06-28
Report Date2018-11-27
Date Reported to FDA2018-11-27
Date Reported to Mfgr2018-11-27
Date Mfgr Received2018-06-04
Device Manufacturer Date2018-01-13
Date Added to Maude2018-06-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SEVERINO
Manufacturer Street861 MARCON BLVD.
Manufacturer CityALLENTOWN PA 18109
Manufacturer CountryUS
Manufacturer Postal18109
Manufacturer Phone4842408332
Manufacturer G1B. BRAUN MELSUNGEN AG
Manufacturer StreetCARL-BRAUN-STR. 1
Manufacturer CityMELSUNGEN, 34212
Manufacturer CountryGM
Manufacturer Postal Code34212
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTROCAN? SAFETY
Generic NameI.V. SAFETY CATHETER
Product CodeDQR
Date Received2018-06-29
Returned To Mfg2018-06-08
Model NumberN/A
Catalog Number4251128-03
Lot Number18A13G8371
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age5 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB. BRAUN MELSUNGEN AG
Manufacturer AddressCARL-BRAUN-STR. 1 MELSUNGEN, HESSEN 34212 GM 34212


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-29