Event Text Entries

[112746964] The basin fluid warmer side pockets became very hot yet the digital display indicated 107 degrees f. The two side "bottle pockets" are not to exceed 110 degrees f. We have found that the bottle temperature can reach 130 degrees f when the bottle is not resting over the temperature sensor. Various size solution bottles may be placed into these pockets. The smaller bottles can be placed in the pocket such that it is not resting over the temperature sensor. When the sensor is not touching the side of the solution bottle it overheats the bottle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7650768
MDR Report Key7650768
Date Received2018-06-29
Date of Report2018-06-05
Date of Event2018-05-31
Report Date2018-06-06
Date Reported to FDA2018-06-06
Date Reported to Mfgr2018-06-29
Date Added to Maude2018-06-29
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTRATEMP
Generic NameWARMER, IRRIGATION SOLUTION
Product CodeLHC
Date Received2018-06-29
Model NumberITW6L
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age4 MO
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK MEDICAL, INC.
Manufacturer Address13000 DEERFIELD PARKWAY, SUITE 300 ALPHARETTA GA 30004 US 30004


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-29