POSEY BED 8070

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-02 for POSEY BED 8070 manufactured by Posey Products Llc.

Event Text Entries

[113282315] Product was received in and analyzed. Evaluation of the returned bed confirmed that one tooth from the left side patient access panel was missing. The malfunction was repaired and returned back to the customer for use. Posey beds are multi-use, serviceable items. As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use. If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing. If a zipper element is missing, it can allow for the unintentional opening of the canopy window. Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper. Although the cause of the missing element cannot be confirmed, it is possible that excessive use or normal wear and tear could have attributed to the damage. Service issues are trended and reviewed by management on a monthly basis. As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted. No corrective or preventative actions are necessary at this time. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[113282316] Customer reported the canopy has a zipper that is split causing the zipper to not fully close. The date tje issue was discovered is not known and no patient or caregiver incident was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020362-2018-00051
MDR Report Key7657290
Date Received2018-07-02
Date of Report2018-06-14
Date Mfgr Received2018-06-14
Device Manufacturer Date2018-03-21
Date Added to Maude2018-07-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactWILLIAM HINCY
Manufacturer StreetPOSEY COMPANY 5635 PECK ROAD
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal91006
Manufacturer Phone6264433143
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NamePOSEY BED 8070
Generic NamePATIENT BED WITH CANOPY/RESTRAINTS
Product CodeOYS
Date Received2018-07-02
Returned To Mfg2018-06-19
Model Number8070
Catalog Number8070
Lot NumberNA
OperatorNURSE
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPOSEY PRODUCTS LLC
Manufacturer Address5635 PECK RD ARCADIA 91006 US 91006


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.