MED HOWMEDICA BONE PLUG 1PK 6215-5-011

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-03 for MED HOWMEDICA BONE PLUG 1PK 6215-5-011 manufactured by Stryker Orthopaedics-mahwah.

Event Text Entries

[113205977] When completed, the evaluation summary will be submitted in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


[113205978] Two separate bone plugs were opened for surgery and deemed unsterile due to seal on packaging. Update 12/jun/2018: spoke to rep. The seal on the outer blister was not adhered to the outer blister all the way around and there were 'gaps' allowing air to pass in and out. The seal on the inner blister was also not adhered to the inner blister all the way around with the same 'gaps'. This happened with 2 devices with 2 different lots. The outer boxes were discarded before the rep could retrieve them. Primary tka procedure. Rep was present in the operating room and witnessed the event. Additional devices were immediately available in the operating room and there was no surgical delay - the surgeon was not even aware of the event until after the procedure was complete.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002249697-2018-02030
MDR Report Key7658868
Date Received2018-07-03
Date of Report2018-10-04
Date of Event2018-06-07
Date Mfgr Received2018-09-06
Device Manufacturer Date2018-03-28
Date Added to Maude2018-07-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOANN LAVATELLI
Manufacturer Street325 CORPORATE DRIVE
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer Phone2018315000
Manufacturer G1STRYKER ORTHOPAEDICS-CORK
Manufacturer StreetIDA INDUSTRIAL ESTATE
Manufacturer CityCARRIGTWOHILL NA
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMED HOWMEDICA BONE PLUG 1PK
Generic NameCEMENT OBTURATOR
Product CodeLZN
Date Received2018-07-03
Returned To Mfg2018-06-14
Catalog Number6215-5-011
Lot NumberCPPUE07BC
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ORTHOPAEDICS-MAHWAH
Manufacturer Address325 CORPORATE DRIVE MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-07-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.