ALERE TOX 9 PANEL INT ICUP DOA-197-251

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional, report with the FDA on 2018-07-03 for ALERE TOX 9 PANEL INT ICUP DOA-197-251 manufactured by Alere San Diego, Inc..

Event Text Entries

[113508648] Investigation conclusion: retention products for the reported lot were tested with in-house drug free donor urine and read at 5 minutes. All devices correctly yielded negative results for all coc150 strips. No false positives were observed. Manufacturing batch record review did not uncover any abnormalities. The presence of cross-reacting substances in the test sample may result in false positive results. See section "cross-reactivity" within the performance characteristics of the package insert for a list of the substances evaluated. Substances not listed have not been evaluated. For this reason, a secondary analytical method must be used to obtain a confirmed result. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[113508649] The customer reported receiving a false positive cocaine result using the alere toxicology icup 9 drug panel for one donor urine specimen as compared to a negative laboratory confirmatory result (lc/ms/ms cut-off=100 ng/ml). The customer reported that only one false positive cocaine result occurred out of 100 devices used. No treatment was provided or withheld based on the false positive result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2018-00082
MDR Report Key7659218
Report SourceCONSUMER,HEALTH PROFESSIONAL,
Date Received2018-07-03
Date of Report2018-07-03
Date Mfgr Received2018-09-17
Date Added to Maude2018-07-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALERE TOX 9 PANEL INT ICUP DOA-197-251
Generic NameENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Product CodeDIO
Date Received2018-07-03
Returned To Mfg2018-07-11
Model NumberDOA-197
Lot NumberDOA7040032
Device Expiration Date2019-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-03