[113456371] The customer contacted siemens customer care center (ccc) and reported discordant, falsely elevated tacrolimus (tac) patient results were obtained on the dimension exl system. Siemens is investigating the event.
Patient Sequence No: 1, Text Type: N, H10


[113456372] Discordant falsely elevated tacrolimus (tac) results were obtained on a patient sample on the dimension exl instrument. The initial results were questioned by the laboratory and not reported to the physician. The same sample was reprocessed on an alternate dimension instrument and an elevated result was obtained. The same sample was also processed with two non-siemens alternate methodologies and lower results were obtained. There was no known impact on patient treatment or care and no alleged patient harm due to the discordant elevated tac results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2018-00402
MDR Report Key7659602
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-07-03
Date of Report2018-10-11
Date of Event2018-05-29
Date Mfgr Received2018-09-11
Device Manufacturer Date2017-12-27
Date Added to Maude2018-07-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer Street500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Street500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal Code197146101
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION?
Generic NameDIMENSION? TAC TACROLIMUS FLEX? REAGENT CARTRIDGE
Product CodeMLM
Date Received2018-07-03
Catalog NumberDF207 SMN10700795
Lot NumberGA8361
Device Expiration Date2018-12-27
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-03