[113626481] (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[113626482] Reverse total shoulder replacement, dr. (b)(6), (b)(6) hospital, (b)(6) 2018. During use on a patient, it was found that one of the four prongs on the locking screw driver main body had broken off and when the surgeon was implanting the first (inferior) locking screw, the screw driver broke the head off that particular screw. In this situation, we had to break the rest of the head off the screw to bury it as we were unable to remove it. A replacement screw driver has since been sent up, however, there is a need for and the surgeon has requested to have a spare screw driver on consignment in any case in the future. No ae to patient: 10 minutes delay in surgery. Patient id: (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[117566180] If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Product complaint #: (b)(4). Investigation summary : the reported event cannot be confirmed due to the product not being returned. Depuy considers the investigation closed. Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1818910-2018-63833
MDR Report Key7665023
Report SourceOTHER
Date Received2018-07-05
Date of Report2018-06-06
Date of Event2018-01-20
Date Mfgr Received2018-08-01
Date Added to Maude2018-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CHAD GIBSON
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465820988
Manufacturer CountryUS
Manufacturer Postal465820988
Manufacturer Phone5743725905
Manufacturer G1DEPUY ORTHOPAEDICS INC 1818910
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465820988
Manufacturer CountryUS
Manufacturer Postal Code465820988
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameUNKNOWN SHOULDER LOCKING SCREW
Generic NameSHOULDER LOCKING SCREW
Product CodeMJT
Date Received2018-07-05
Catalog NumberUNK SHOULDER LOCKING SCREW
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY ORTHOPAEDICS INC 1818910
Manufacturer Address700 ORTHOPAEDIC DRIVE WARSAW IN 465820988 US 465820988


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.