ANIMAS VIBE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-05 for ANIMAS VIBE manufactured by Animas Llc.

Event Text Entries

[113525906] The pump has not been returned to animas for evaluation. If the device is returned, an evaluation shall be completed and a supplemental report will be filed. No conclusions can be made at this time.
Patient Sequence No: 1, Text Type: N, H10


[113525907] On (b)(6) 2018, the reporter contacted animas, alleging a display (line through display with moisture) issue. It was alleged there was moisture behind the display. There was no indication that the device caused or contributed to an adverse event. This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
Patient Sequence No: 1, Text Type: D, B5


[117023682] Device evaluation: the device has been returned and evaluated by product analysis on 17-jul-2018 with the following findings: during investigation, the pump powered up to a fully functional display. The alleged line through the display was unable to be duplicated. Visible moisture was found in the battery compartment with a leak. The pump was opened to find moisture corrosion on the battery housing. No moisture was found on the display.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2531779-2018-12366
MDR Report Key7665497
Date Received2018-07-05
Date of Report2018-06-27
Date Mfgr Received2018-06-27
Date Added to Maude2018-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARIN SARGRAD
Manufacturer Street200 LAWRENCE DR
Manufacturer CityWEST CHESTER PA 193803428
Manufacturer CountryUS
Manufacturer Postal193803428
Manufacturer Phone4843561808
Manufacturer G1ANIMAS LLC
Manufacturer Street200 LAWRENCE DR
Manufacturer CityWEST CHESTER PA 193803428
Manufacturer CountryUS
Manufacturer Postal Code193803428
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameANIMAS VIBE
Generic NameINSULIN INFUSION PUMP
Product CodePOY
Date Received2018-07-05
Returned To Mfg2018-07-10
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerANIMAS LLC
Manufacturer Address200 LAWRENCE DR WEST CHESTER PA 193803428 US 193803428


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-05