VIDAS? CMV IGM 30205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-07-06 for VIDAS? CMV IGM 30205 manufactured by Biomerieux Sa.

Event Text Entries

[113339898] A customer from (b)(6) notified biom? Rieux of obtaining equivocal and false negative results when testing their internal quality control (iqc) virotrol? Torch-m sample with a new batch of vidas? Cmv igm, lot 1005897340. In spite of the discrepant results, the new batch was implemented 21-oct-2017. The iqc functioned as intended by sending a warning to the customer. The customer checked the results and found no impact to patient samples. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There was no patient associated with this discrepant quality control result. Although product reference 30205 is not sold/distributed in the united states, a similar product is (reference 30205-01). A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00098
MDR Report Key7666722
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-07-06
Date of Report2018-09-19
Date Mfgr Received2018-08-24
Device Manufacturer Date2017-08-11
Date Added to Maude2018-07-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. CANDACE MARTIN
Manufacturer Street595 ANGLUM ROAD
Manufacturer CitySAINT LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX SA
Manufacturer Street376 CHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? CMV IGM
Generic NameVIDAS? CMV IGM
Product CodeLKQ
Date Received2018-07-06
Model Number30205
Device Expiration Date2018-06-21
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address376 CHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.