NEOPROBE NPB09S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-07-09 for NEOPROBE NPB09S manufactured by Devicor Medical Products Inc.

Event Text Entries

[113510904] The neoprobe npb09s is a reusable gamma detection probe that is cleaned and disinfected and/or sterilized between uses. Based on the review of the dhr, the device was manufactured in oct. 2013, and released for commercial use having met all acceptance tests and calibration criteria. The device has not been received at the site of manufacture, nortech. A supplemental report will be submitted once the evaluation is complete. While there was no adverse effect during this event, there is a potential for serious injury if malfunction were to reoccur. This potential serious injury would be from an infection due to cross contamination and/or the tip of the neoprobe falling off during the procedure and being left behind in the body. Possible causes of this defect would be deficient expoxy or a loose probe, pursuant to 21 cfr 803, we are deeming this a reportable event and thus submitting this medwatch report.
Patient Sequence No: 1, Text Type: N, H10


[113510905] It was reported by the sales rep that after the procedure the probe has a broken tip. The procedure was completed with device with no patient consequence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008492462-2018-00049
MDR Report Key7671228
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-07-09
Date of Report2018-07-07
Date of Event2018-06-13
Date Mfgr Received2018-06-13
Device Manufacturer Date2013-10-04
Date Added to Maude2018-07-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS SHAWNA ROSE
Manufacturer Street300 E-BUSINESS WAY FIFTH FLOOR
Manufacturer CityCINCINNATI OH 45241
Manufacturer CountryUS
Manufacturer Postal45241
Manufacturer Phone5138649178
Manufacturer G1NORTECH SYSTEMS, INC.
Manufacturer Street925 SIXTH AVENUE, NE
Manufacturer CityMILACA MN 56353
Manufacturer CountryUS
Manufacturer Postal Code56353
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEOPROBE
Generic NamePROBE, UPTAKE, NUCLEAR
Product CodeIZD
Date Received2018-07-09
Model NumberNPB09S
Catalog NumberNPB09S
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDEVICOR MEDICAL PRODUCTS INC
Manufacturer Address300 E-BUSINESS WAY FIFTH FLOOR CINCINNATI OH 45241 US 45241


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.