VIDAS? C. DIFFICILE GDH 30125

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-07-10 for VIDAS? C. DIFFICILE GDH 30125 manufactured by Biomerieux Sa.

Event Text Entries

[113910850] A customer in (b)(6) reported to biom? Rieux that they obtained a false positive result when using the product vidas? Gdh (ref. 30125), lot 1006161620. The customer performed a correlation study between two vidas 3 analyzers (vidas (b)(4) and vidas (b)(4)) for vidas gdh. The results were obtained as follows: sample 1: stool volume not doubled: 1st test on vidas (b)(4) was obtained positive at 0. 25, 2nd test on vidas (b)(4) was obtained negative at 0. 00. Sample 2: doubled stool volume and volume of r1: 1st test on vidas (b)(4) was obtained negative at 0. 00 (non-standard c1 control), 2nd test on vidas (b)(4) was obtained negative at 0. 00 (non-standard c1 control). There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00104
MDR Report Key7674193
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-07-10
Date of Report2019-01-08
Date Mfgr Received2018-12-13
Device Manufacturer Date2017-12-14
Date Added to Maude2018-07-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JEFF SCANLAN
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318694
Manufacturer G1BIOMERIEUX SA
Manufacturer Street376, CHEMIN DE L'ORME
Manufacturer CityMARCY L'ETOILE, 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? C. DIFFICILE GDH
Generic NameVIDAS? GDH
Product CodeMCB
Date Received2018-07-10
Catalog Number30125
Lot Number1006161620
Device Expiration Date2019-01-27
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address376, CHEMIN DE L'ORME MARCY L'ETOILE, 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-10