MST MALYUGIN RING SYSTEM MAL-1002-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-07-10 for MST MALYUGIN RING SYSTEM MAL-1002-1 manufactured by Microsurgical Technology, Inc..

Event Text Entries

[113568503] The ring was removed from the eye, a second malyugin ring assembly was used and surgery was completed. After the device was returned, the malyugin ring assembly passed functional testing with a new ring. The ring used in surgery was deformed and could not be evaluated. Manufacturing records were pulled and the ring was within specification during inspection.
Patient Sequence No: 1, Text Type: N, H10


[113568504] The customer reported that upon insertion of the malyugin ring, the proximal scroll would not eject fully from the cannula when in the eye. The scroll caught the iris and tore a small fragment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2018-00022
MDR Report Key7674652
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-07-10
Date of Report2018-07-10
Date Mfgr Received2018-06-12
Device Manufacturer Date2018-03-23
Date Added to Maude2018-07-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CECIL CHRISINGER
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer G1MICROSURGICAL TECHNOLOGY, INC.
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMST MALYUGIN RING SYSTEM
Generic NameCLIP IRIS RETRACTOR
Product CodeHOC
Date Received2018-07-10
Returned To Mfg2018-06-28
Model NumberMAL-1002-1
Catalog NumberMAL-1002-1
Lot Number097119
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY, INC.
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052