IP ACTIVFLO BIOPSY III, ORANGE TISSUE CASSETTES 39LC-625-7

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-11 for IP ACTIVFLO BIOPSY III, ORANGE TISSUE CASSETTES 39LC-625-7 manufactured by Leica Biosystems Richmond.

Event Text Entries

[113740988] As per customer contact, one patient ((b)(6 )year old male) specimen (rectum polyp) in 1 leica activflo biopsy iii cassette was loaded on a vip processor on friday, (b)(6) 2018 to come off on monday, (b)(6) at 5:30 am. The lid of the cassette reportedly opened during processing and the specimen was lost. Total number of cassettes loaded in the run was 71. A sample of the cassettes was provided to the sales representative by the customer and was sent back to the manufacturer for evaluation. Additionally, product of the same lot reported was pulled from inventory for testing by qa histotechs to attempt to replicate customer complaint. All cassettes remained closed during processing. Upon the receipt of product returned by the customer, the product was sent to qa histotechs to attempt to replicate customer complaint. All cassettes remained closed. In opening the cassettes after processing, there was the expected amount of resistance in removing the lids. All cassettes were found to function as designed.
Patient Sequence No: 1, Text Type: N, H10


[113740989] On monday, (b)(6), a sales representative of leica biosystems was notified of patient tissue lost by the customer due to the cassettes reportedly opening during processing. Patient must undergo another biopsy due to tissue loss.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419341-2018-00003
MDR Report Key7679701
Date Received2018-07-11
Date of Report2018-07-11
Date of Event2018-06-11
Date Mfgr Received2018-06-11
Date Added to Maude2018-07-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PETER SPARACIO
Manufacturer Street5205 US HIGHWAY 12
Manufacturer CityRICHMOND IL 60071
Manufacturer CountryUS
Manufacturer Postal60071
Manufacturer Phone8156782000
Manufacturer G1LEICA BIOSYSTEMS RICHMOND
Manufacturer Street5205 US HIGHWAY 12
Manufacturer CityRICHMOND IL 60071
Manufacturer CountryUS
Manufacturer Postal Code60071
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameIP ACTIVFLO BIOPSY III, ORANGE TISSUE CASSETTES
Generic NameTISSUE CASSETTES
Product CodeIDZ
Date Received2018-07-11
Model Number39LC-625-7
Lot Number625-LID-7-171221-A Q14
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS RICHMOND
Manufacturer Address5205 US HIGHWAY 12 RICHMOND IL 60071 US 60071


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-07-11