[16166372] A dentist reported that a pt received a burn to his tongue during the use of a 25lpa handpiece. The pt was prescribed an antibiotic and tylenol. It was reported that the pt has since recovered.
Patient Sequence No: 1, Text Type: D, B5


[16200606] The actual unit associated with the incident was returned to the mfr for analysis. Visual inspection of the unit revealed that the head was dented at the back cap and in several places around the water spray plate. The bearings were noted to grind and heat up upon operation of the unit. The unit was disassembled for further analysis and light residue inside the unit was observed. A cautionary statement is included with each unit, which states that electric micromotors generate significantly more power than traditional air turbines and air motors. Due to increased torque and speed, poorly maintained, damaged or misused handpieces can potentially generate frictional heat capable of causing serious burn injuries to the pt. It also states that the head of the attachment should not be used as a cheek or tongue retractor.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1419798-2006-00004
MDR Report Key768383
Report Source05,07
Date Received2006-10-06
Date of Report2006-09-08
Date of Event2006-08-31
Date Mfgr Received2006-09-08
Device Manufacturer Date2005-01-01
Date Added to Maude2006-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactJOHN MILLER, DIR
Manufacturer Street901 WEST OAKTON STREET
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone8473643931
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELECTROTORQUE PLUS
Generic NameELECTRIC DENTAL HANDPIECE
Product CodeEKK
Date Received2006-10-06
Returned To Mfg2006-09-14
Model Number25LPA
Catalog Number1002.1526
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key756255
ManufacturerKAVO DENTAL
Manufacturer Address340 EAST MAIN ST. LAKE ZURICH IL 60047 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-10-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.