HOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE DFH-0008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-12 for HOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE DFH-0008 manufactured by Microsurgical Technology, Inc..

Event Text Entries

[114056024] The device evaluation showed that the blade was broken off from the top of the tine, the other blade broke at the base of the tine. There was moderate corrosion at the site of the fractures and throughout the cannula of the scissors. Small rust spots visible and corrosion pits. The mouth of the cannula is deformed, flared where the tines had made contact, indicating that scissors came into contact with hard object. Device failure is due to corrosion and inadequate use, these scissors are not made to cut the iol due to how hard the iol is.
Patient Sequence No: 1, Text Type: N, H10


[114056025] The customer reported the blade was caught in the iol and fell off in the eye during cataract surgery. The blade and iol was removed during surgery and no known harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2018-00024
MDR Report Key7684111
Date Received2018-07-12
Date of Report2018-07-12
Device Manufacturer Date2012-02-29
Date Added to Maude2018-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CECIL CHRISINGER
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer G1MICROSURGICAL TECHNOLOGY, INC.
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE
Generic NameOPHTHALMIC SCISSORS
Product CodeHNF
Date Received2018-07-12
Returned To Mfg2018-07-07
Model NumberDFH-0008
Catalog NumberDFH-0008
Lot Number045617
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY, INC.
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-07-12