RAYSTATION 6.0

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-13 for RAYSTATION 6.0 manufactured by Raysearch Laboratories Ab (publ).

Event Text Entries

[114540691] User reported a significant dose hotspot in the inferior corner of an off-axis field. The deviation was detected during treatment planning and had no impact on the patient. The root cause was that the off-axis beam profile correction in the beam model had very high values beyond 20 cm. Clinic had failed to validate the beam model for very large or off-axis fields. They received assistance in adjusting the beam model. This was abnormal use, using the system contrary to instructions. The responsibility of the user to validate the beam model for all clinical situations for which it is to be used is clearly stated in the ifu, giving field size and off-axis fields as explicit examples of what needs to be included. The clinic's responsibility to validate the beam model is also "common knowledge" and has not been disputed by the customers. However, since more than one clinic has been found to use similar, incorrect beam models, a fsn has been issued (similar incident previously reported in 3007774465-2018-00002).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007774465-2018-00004
MDR Report Key7687591
Date Received2018-07-13
Date of Report2018-07-13
Date of Event2018-04-25
Date Mfgr Received2018-04-25
Device Manufacturer Date2016-12-14
Date Added to Maude2018-07-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. DAVID HEDFORS
Manufacturer StreetPO BOX 3297
Manufacturer CitySTOCKHOLM, 10365
Manufacturer CountrySW
Manufacturer Postal10365
Manufacturer G1RAYSEARCH LABORATORIES AB (PUBL)
Manufacturer StreetSVEAVAGEN 44
Manufacturer CitySTOCKHOLM, 11134
Manufacturer CountrySW
Manufacturer Postal Code11134
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameRAYSTATION
Generic NameRADIATION THERAPY TREATMENT PLANNING SYSTEM
Product CodeMUJ
Date Received2018-07-13
Model Number6.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRAYSEARCH LABORATORIES AB (PUBL)
Manufacturer AddressSVEAVAGEN 44 STOCKHOLM, 11134 SW 11134


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-07-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.