VIDAS? VARICEL ZOSTER IGG 30217

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-07-17 for VIDAS? VARICEL ZOSTER IGG 30217 manufactured by Biomerieux Sa.

Event Text Entries

[114835671] A customer in (b)(6) notified biom? Rieux of discrepant results associated with vidas? Varicel zoster igg (reference 30217). The customer reported during the last (b)(6) distribution for immunity screen scheme participants using biom? Rieux vidas assay reported equivocal or negative vzv igg results for specimen (b)(4) that was prepared from a single neat donation and was positive for vzv igg. In pre-distribution testing the specimen was tested positive using diasorin liaison, virion serion, binding site vacczyme quantitative assays. Specimen 4372 was tested equivocal using biom? Rieux vidas? Assay. Readings from pre-distribution and from participants' results show that vzv igg antibodies were present in the specimen. Below references pre-distribution test results for specimen (b)(4): biom? Rieux vidas vzg tv 0. 74 equivocal; diasorin 516. 6 miu/ml (positive cut off 100;) virion serion vzv igg : 300. 58 iu/ml (grey zone 15. 0-20. 0); bonding site vacczyme vzv glycoprotein igg 207 miu/ml (grey zone 100-150). Out of the 93 biom? Rieux vidas? Users 84 (90. 3%) reported an equivocal result, 7 (7. 5%) reported a negative result and one (2. 2%) user reported a positive result for specimen (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00107
MDR Report Key7694066
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-07-17
Date of Report2018-12-11
Date Mfgr Received2018-11-13
Date Added to Maude2018-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIFFANY HALL
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SA
Manufacturer Street376 CHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? VARICEL ZOSTER IGG
Generic NameVIDAS? VARICEL ZOSTER IGG
Product CodeLFY
Date Received2018-07-17
Catalog Number30217
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address376 CHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.