ALERE BINAXNOW STREPTOCOCCUS PNEUMONIAE ANTIGEN CARD 710-000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-07-18 for ALERE BINAXNOW STREPTOCOCCUS PNEUMONIAE ANTIGEN CARD 710-000 manufactured by Alere Scarborough Inc..

Event Text Entries

[114817419] Testing was performed at alere (b)(4) on retained device lot e04180 associated with kit lot e04568 with a low positive and a negative control. All test results were valid and performed as expected. Additionally, the manufacturing batch records for lot e04180 were reviewed. This lot met the required release specifications. A review of the complaints reported as false negative related to lot number e04180 showed no similar complaints. The evidence available does not indicate that the product is performing outside label claims. Alere (b)(4) was unable to determine the exact root cause of the reported issue. The available evidence suggests that this device lot is performing within labeled claims.
Patient Sequence No: 1, Text Type: N, H10


[114817420] This report represent 2 of at least three 3 false negative alere binaxnow streptococcus pneumoniae antigen card patient test results reported by a customer. Cultures were reported positive and one of the samples was also positive by a fluorescence method. The exact number of patients involved is unknown. The sample type and patient(s) outcome are unknown. There is insufficient information to determine if a malfunction occurred. In a worst-case scenario, a reported false negative result occurred and a cerebral spinal fluid (csf) may have been used. A false negative result may cause a delay in the patient receiving appropriate therapy. As meningitis is a serious, life threatening condition delays in treatment may have severe consequences; therefore, this is a reportable event. This mdr is a retrospective filing in response to an observation from an fda inspection conducted february 26th to march 2nd , 2018 at alere (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1221359-2018-00477
MDR Report Key7699619
Report SourceHEALTH PROFESSIONAL
Date Received2018-07-18
Date of Report2018-07-18
Date of Event2016-06-20
Date Mfgr Received2018-06-20
Device Manufacturer Date2016-01-05
Date Added to Maude2018-07-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBENJAMIN CRYSTAL
Manufacturer Street10 SOUTHGATE ROAD
Manufacturer CitySCARBOROUGH ME 04074
Manufacturer CountryUS
Manufacturer Postal04074
Manufacturer Phone2077305820
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALERE BINAXNOW STREPTOCOCCUS PNEUMONIAE ANTIGEN CARD
Generic NameANTISERA, ALL GROUPS, STREPTOCOCCUS SPP
Product CodeGTZ
Date Received2018-07-18
Catalog Number710-000
Lot NumberE04568
Device Expiration Date2017-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SCARBOROUGH INC.
Manufacturer Address10 SOUTHGATE ROAD SCARBOROUGH ME 04074 US 04074


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.