ILED 7 4047020

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-19 for ILED 7 4047020 manufactured by Trumpf Medizin Systeme Gmbh + Co. Kg.

Event Text Entries

[114801341] The lamp head is secured to the spring arm by a retaining segment which is maintained in place by a safety sleeve that is held over the segment with a set screw. The root cause for this incident has been determined as a missing set screw that allowed the safety sleeve to move out of place and the retaining segment to migrate out. The service history of the device was reviewed and revealed no servicing activity performed by trumpf medical that would have any impact on this connection. The account's biomedical staff reassembled the lamp head after the fall and a trumpf medical service technician completed an inspection and functions check of the system the same day. All of the facility's light systems have been evaluated for the same condition and no other issues were identified.
Patient Sequence No: 1, Text Type: N, H10


[114801342] As the hospital staff was cleaning the light system, the lamp head of a trumpf medical surgical light separated at the cardanic - spring arm joint then fell. The lamp head dropped a short distance landing directly on an or table top. No one was injured as a result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681407-2018-00035
MDR Report Key7702440
Date Received2018-07-19
Date of Report2018-06-22
Date of Event2018-06-21
Date Mfgr Received2018-06-22
Device Manufacturer Date2017-05-18
Date Added to Maude2018-07-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFFEN ULBRICH
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer CitySAALFELD, 07318
Manufacturer CountryGM
Manufacturer Postal07318
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameILED 7
Generic NameSURGICAL LIGHT
Product CodeFQP
Date Received2018-07-19
Model Number4047020
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer AddressCARL-ZEISS STRASSE 7-9 SAALFELD, 07318 GM 07318


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-19

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.