VIDAS? CMV IGM 30205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-07-19 for VIDAS? CMV IGM 30205 manufactured by Biomerieux Sa.

Event Text Entries

[114823476] A customer in (b)(6) notified biom? Rieux of discrepant results associated with vidas? Cmv igm (reference (b)(4)). The customer reported running the test on abbott architect and got a positive igm result (1. 15 au/ml) and a positive igg result (200 au/ml). The results were also run on vidas? As a second method and the results were negative (0. 55 au/ml) igm and positive igg (26 au/ml), and a low avidity result of 0. 26 au/ml. Patient results were impacted. It is unknown if the result was reported to the treating physician. There was a delay in reporting results due to this discrepancy. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00111
MDR Report Key7702923
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-07-19
Date of Report2018-09-26
Date Mfgr Received2018-08-30
Device Manufacturer Date2018-02-19
Date Added to Maude2018-07-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIFFANY HALL
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SA
Manufacturer Street376 CHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? CMV IGM
Generic NameVIDAS? CMV IGM
Product CodeLKQ
Date Received2018-07-19
Catalog Number30205
Lot Number1006302830
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address376 CHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-19