ARCHITECT PROLACTIN 07K76-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-07-19 for ARCHITECT PROLACTIN 07K76-25 manufactured by A.i.d.d Longford.

Event Text Entries

[114810301] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. Patient information, no further patient information was provided.
Patient Sequence No: 1, Text Type: N, H10


[114810302] The customer reported a falsely depressed architect prolactin result for a patient with a large pituitary tumor (prolactinoma). Sample id (b)(4) generated a result of 106. 5 miu/l which the customer believed was lower than expected based on the patient's clinical picture. The customer further tested the sample with multiple dilutions including 1:10 autodilution and manual dilutions with multi-assay diluent as recommended by the package insert. All dilutions generated results below the reportable limit. The sample was further evaluated with multiple dilutions using a serum sample with a low prolactin (22. 6 miu/l). The results ranged from 125. 8 miu/l with a 1:2 dilution to 28701. 6 miu/l with a 1:1200 dilution. No adverse impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005094123-2018-00024
MDR Report Key7702977
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-07-19
Date of Report2018-08-27
Date of Event2018-07-01
Date Mfgr Received2018-08-22
Device Manufacturer Date2018-01-23
Date Added to Maude2018-07-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1A.I.D.D LONGFORD
Manufacturer StreetLISNAMUCK CO. LONGFORD
Manufacturer CityLONGFORD NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT PROLACTIN
Generic NamePROLACTIN
Product CodeCFT
Date Received2018-07-19
Catalog Number07K76-25
Lot Number84028UI00
Device Expiration Date2018-11-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerA.I.D.D LONGFORD
Manufacturer AddressLISNAMUCK CO. LONGFORD LONGFORD NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.