Event Text Entries

[114806252] This type of product performance issue has not been reported previously. During the on-site incident investigation, several screws that connect the chassis motor to the frame were found to be loose or missing. The unit was sent to harmar for further investigation in may where it was determined that the screws were loose or missing due to the lack of loctite threadlocker application on the screws.
Patient Sequence No: 1, Text Type: N, H10


[114806253] On (b)(6) 2018 the user got on the stair lift chair at the bottom step and was riding upwards to the top landing when the stair lift made a popping sound and the stair lift seat/chassis descended back down the staircase to the end of the rail. The user stepped off the stair lift as it descended and sustained minor pain in her leg and a small cut on her finger.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004153586-2018-00001
MDR Report Key7705796
Date Received2018-07-20
Date of Report2018-07-17
Date of Event2018-05-05
Date Mfgr Received2018-05-08
Device Manufacturer Date2018-04-23
Date Added to Maude2018-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DAWN WATTERS
Manufacturer Street2075 47TH STREET
Manufacturer CitySARASOTA FL 34234
Manufacturer CountryUS
Manufacturer Postal34234
Manufacturer Phone9413087398
Manufacturer G1HARMAR MOBILITY
Manufacturer Street18505 E 163RD STREET
Manufacturer CityLAKE WINNEBAGO MO 64034
Manufacturer CountryUS
Manufacturer Postal Code64034
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePINNACLE HEAVY DUTY
Generic NamePOWERED STAIRWAY CHAIR LIFT
Product CodePCD
Date Received2018-07-20
Returned To Mfg2018-05-10
Model NumberSL600HD
Catalog NumberSL600HD
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHARMAR MOBILITY, LLC
Manufacturer Address18505 E 163RD STREET LAKE WINNEBAGO MO 64034 US 64034


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-20