MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-07-23 for DIMENSION? DF207 SMN 10700795 manufactured by Siemens Healthcare Diagnostics Inc..
[114693430]
Mdrs 2517506-2018-00422, 2517506-2018-00423, 2517506-2018-00424, 2517506-2018-00425, were filed for the same event. The customer contacted siemens customer care center (ccc) and reported a discordant, falsely elevated tacrolimus (tac) patient result was obtained on the dimension exl system. Siemens is investigating the event.
Patient Sequence No: 1, Text Type: N, H10
[114693431]
A discordant falsely elevated tacrolimus (tac) result was obtained on a patient sample on the dimension exl instrument. The same sample was processed on an alternate methodology and a lower result was obtained. There was no known impact on patient treatment or care and no alleged patient harm due to the discordant elevated tac result.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2517506-2018-00426 |
MDR Report Key | 7710153 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2018-07-23 |
Date of Report | 2018-11-09 |
Date of Event | 2018-06-23 |
Date Mfgr Received | 2018-10-29 |
Device Manufacturer Date | 2018-04-20 |
Date Added to Maude | 2018-07-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | JAMES MORGERA |
Manufacturer Street | 500 GBC DRIVE PO BOX 6101 |
Manufacturer City | NEWARK DE 197146101 |
Manufacturer Country | US |
Manufacturer Postal | 197146101 |
Manufacturer Phone | 3026318356 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Street | 500 GBC DRIVE PO BOX 6101 |
Manufacturer City | NEWARK DE 197146101 |
Manufacturer Country | US |
Manufacturer Postal Code | 197146101 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DIMENSION? |
Generic Name | DIMENSION? TAC TACROLIMUS FLEX? REAGENT CARTRIDGE |
Product Code | MLM |
Date Received | 2018-07-23 |
Catalog Number | DF207 SMN 10700795 |
Lot Number | GA9106 |
Device Expiration Date | 2019-04-16 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Address | 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-07-23 |