CLINICAL CHEMISTRY CALCIUM 03L79-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-07-23 for CLINICAL CHEMISTRY CALCIUM 03L79-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[114725742] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. There is no further patient information provided by customer due to privacy issues.
Patient Sequence No: 1, Text Type: N, H10


[114725743] The customer reported falsely elevated calcium results on one patient. The initial sample was drawn at the physicians site: sample from (b)(6) 2018 (b)(6) on (b)(6) 2018 ca = 4. 26 / 4. 16 / 4. 05 / 3. 9 mmol/l(2. 23-2. 5 mmol/l). The results were reported and the patient was sent to the emergency room where a new sample was drawn: (b)(6) ca = 2. 37. The original (b)(6) was repeated on (b)(6) 2018 ca = 2. 4; on (b)(6) 2018 different tube same draw (b)(6) ca run 21x and the results ranged from 1. 91 to 2. 06. There was no reported impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2018-00362
MDR Report Key7710691
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-07-23
Date of Report2018-09-10
Date of Event2018-07-05
Date Mfgr Received2018-09-07
Device Manufacturer Date2018-04-01
Date Added to Maude2018-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY CALCIUM
Generic NameCALCIUM
Product CodeCJY
Date Received2018-07-23
Catalog Number03L79-21
Lot Number98314UN17
Device Expiration Date2019-02-27
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.