PRISM CHAGAS 07K35-68

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-07-23 for PRISM CHAGAS 07K35-68 manufactured by Abbott Laboratories.

Event Text Entries

[114819810] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. An evaluation is in process.
Patient Sequence No: 1, Text Type: N, H10


[114819811] The customer observed multiple falsely reactive results while using the prism chagas reagents. The following data was provided. Sid numbers were provided and the customer stated repeat reactive specimens did not confirm (confirmatory testing negative), however specific results values were not provided. (b)(6). No impact to patient or donor management were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1415939-2018-00114
MDR Report Key7711415
Report SourceHEALTH PROFESSIONAL
Date Received2018-07-23
Date of Report2018-10-29
Date of Event2018-05-01
Date Mfgr Received2018-10-19
Device Manufacturer Date2018-02-01
Date Added to Maude2018-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600643500
Manufacturer CountryUS
Manufacturer Postal Code600643500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRISM CHAGAS
Generic NameCHAGAS
Product CodeMIU
Date Received2018-07-23
Catalog Number07K35-68
Lot Number85017M500
Device Expiration Date2018-12-13
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US 600643500


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.