MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-07-23 for ENDO GRASP 173030 manufactured by Us Surgical Puerto Rico.
[114934703]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[114934704]
According to the reporter, during laparoscopic choledocholithotomy procedure, the jaw was broken on the roots while clamping tissue. The part fell into the patient's cavity and was retrieved. The tissue was not so thick and no excess force was not applied to the device. The procedure was completed with another device. There was no patient injury.
Patient Sequence No: 1, Text Type: D, B5
[122530293]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[126162169]
Evaluation summary: post market vigilance (pmv) led an evaluation of one device. One jaw of the device was broken at the fulcrum. Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture. The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition. Replication of the reported condition may occur when the device is exposed to a side force (leverage) that consequently breaks one side of the jaws. Another possibility is when the device is activated with too much force to the jaws and is used in a twisting motion resulting in breakage. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2647580-2018-03668 |
MDR Report Key | 7711532 |
Date Received | 2018-07-23 |
Date of Report | 2018-11-02 |
Date of Event | 2018-06-28 |
Date Mfgr Received | 2018-10-09 |
Device Manufacturer Date | 2017-08-09 |
Date Added to Maude | 2018-07-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LISA HERNANDEZ |
Manufacturer Street | 60 MIDDLETOWN AVE. |
Manufacturer City | NORTH HAVEN CT 06473 |
Manufacturer Country | US |
Manufacturer Postal | 06473 |
Manufacturer Phone | 2034925563 |
Manufacturer G1 | US SURGICAL PUERTO RICO |
Manufacturer Street | 201 SABANETAS INDUSTRIAL PARK |
Manufacturer City | PONCE PR 007164401 |
Manufacturer Country | US |
Manufacturer Postal Code | 007164401 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | ENDO GRASP |
Generic Name | FORCEPS, OBSTETRICAL |
Product Code | HDA |
Date Received | 2018-07-23 |
Returned To Mfg | 2018-07-13 |
Model Number | 173030 |
Catalog Number | 173030 |
Lot Number | P7H1458PX |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | US SURGICAL PUERTO RICO |
Manufacturer Address | 201 SABANETAS INDUSTRIAL PARK PONCE PR 007164401 US 007164401 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-07-23 |