[114946653] A corindus service engineer inspected the touchscreen and determined that it contained particles from an ammonium salt based cleaning solution. The corpath grx operator's manual states that "isopropyl alcohol (ipa) based disinfectants, germicides such as spor-klenz, or 10% to 20% bleach solutions" are recommended for cleaning. The use of an alternate cleaning solution may have led to the failure of the robotic drive touchscreen. The service engineer replaced the touchscreen and system is now functioning as intended.
Patient Sequence No: 1, Text Type: N, H10


[114946654] The robotic drive touchscreen became unresponsive when the technician selected "load/exchange device" to remove the guidewire. The system was rebooted but the "continue case" button could not be selected due to the unresponsive touchscreen. The emergency release button on the cassette had to be pressed to disassemble the cassette and remove the interventional devices from the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007822508-2018-00005
MDR Report Key7713953
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-07-24
Date of Report2018-07-24
Date of Event2018-07-09
Date Mfgr Received2018-07-09
Device Manufacturer Date2017-03-01
Date Added to Maude2018-07-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT LAVADO
Manufacturer Street309 WAVERLEY OAKS ROAD SUITE 105
Manufacturer CityWALTHAM MA 02452
Manufacturer CountryUS
Manufacturer Postal02452
Manufacturer Phone5086533335
Manufacturer G1CORINDUS, INC.
Manufacturer Street309 WAVERLEY OAKS ROAD SUITE 105
Manufacturer CityWALTHAM MA 02452
Manufacturer CountryUS
Manufacturer Postal Code02452
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCORPATH GRX SYSTEM
Generic NameCORPATH GRX SYSTEM
Product CodeDXX
Date Received2018-07-24
Model Number303
Catalog Number303
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCORINDUS, INC.
Manufacturer Address309 WAVERLEY OAKS ROAD SUITE 105 WALTHAM MA 02452 US 02452


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.