LIFECODES HLA-DQA1/B1 SSO TYPING KIT 628930

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-07-24 for LIFECODES HLA-DQA1/B1 SSO TYPING KIT 628930 manufactured by Immucor Gti Diagnostics, Inc..

Event Text Entries

[114943405] Customer contacted immucor technical support (complaint #(b)(4)) that samples on the lifecodes hla-dqa1/b1 sso typing kit (628930), lot # 3004970 had discrepant results. Sample was resulting as dqb1*06:09:01:01 homozygous. It was subsequently identified that the sample was not homozygous but found to be dqb1 *06:09:01:01, *03:03:02:01. The customer was not following the product instructions for use (ifu). The customer was not using full volumes of reagents. The customer identified the issue because the obtained results did not fit with the hla-dr associations they should match. The customer repeated the testing with an alternate method (ssp) and discovered a hla-dq6, dq3 result rather than the hla-dq6 homozygote. The customer informed immucor that they performed an investigation to check for patient harm but have found no cause for patient harm. Customer is not following the ifu and was aware of potential risks involved in doing this.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183608-2018-00023
MDR Report Key7716242
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-07-24
Date of Report2018-03-02
Date of Event2018-02-09
Date Mfgr Received2018-03-02
Device Manufacturer Date2017-04-20
Date Added to Maude2018-07-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS JULIE EWEND
Manufacturer Street20925 CROSSROADS CIRCLE
Manufacturer CityWAUKESHA WI 53186
Manufacturer CountryUS
Manufacturer Postal53186
Manufacturer Phone2627541009
Manufacturer G1IMMUCOR GTI DIAGNOSTICS, INC.
Manufacturer Street20925 CROSSROADS CIRCLE
Manufacturer CityWAUKESHA WI 53186
Manufacturer CountryUS
Manufacturer Postal Code53186
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIFECODES HLA-DQA1/B1 SSO TYPING KIT
Generic NameLIFECODES HLA-DQA1/B1 SSO TYPING KIT
Product CodeMZI
Date Received2018-07-24
Catalog Number628930
Lot Number3004970
Device Expiration Date2018-06-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR GTI DIAGNOSTICS, INC.
Manufacturer Address20925 CROSSROADS CIRCLE WAUKESHA WI 53186 US 53186


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-24