MECHANICAL CHAIR/TRANSPORT CHAIR UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-07-26 for MECHANICAL CHAIR/TRANSPORT CHAIR UNKNOWN manufactured by Champion Manufacturing Inc.

Event Text Entries

[115123206] Assuming this is a champion chair (which could not be verified), it has been determined this is an operational issue. Review of operator manual provides proper operation of foot tray and warnings on page 14. Review of the history of this facility shows they have purchased over 100 chairs on a periodic basis from champion from 2004 through 2017. Review of service parts and sales activity over that timeframe shows no repairs done to their chairs or mechanical parts purchased that would pertain to any foot trays. Facility ordered (140) sets of 3" casters june 4, 2018.
Patient Sequence No: 1, Text Type: N, H10


[115123207] On 4/6/2018 phone call from facility was to order replacement parts. The facility wanted to order optional smaller casters. There was no complaint stated, but it was mentioned that the facility decided to remove the foot trays due to patient falls and now the chairs are too high for many patients. Facility was made aware of the optional yellow foot tray and with a caution label in english, french and spanish for higher visibility, they declined that option. Caution label states: caution weight capacity 225 lbs. Stow when not in use. Use with staff assistance only. On 5/31/2018 phone call advised that sometime near the beginning of the year (2018) a patient sitting in a chair, she stood up unattended on the foot tray that was left extended, took a step and fell. The patient fell and fractured her femur, which required surgery to repair. On 6/15/2018 phone call produced the information regarding the individual involved and information that no serial number was recorded for this incident, no date or time of incident was reported. It is assumed this is a champion product but that cannot be verified.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1834066-2018-00003
MDR Report Key7722751
Report SourceUSER FACILITY
Date Received2018-07-26
Date of Report2018-07-24
Date Mfgr Received2018-06-15
Date Added to Maude2018-07-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTINE MILLER
Manufacturer Street2601 INDUSTRIAL PARKWAY
Manufacturer CityELKHART IN 46516
Manufacturer CountryUS
Manufacturer Postal46516
Manufacturer Phone8009985018
Manufacturer G1CHAMPION MANUFACTURING INC
Manufacturer Street2601 INDUSTRIAL PARKWAY
Manufacturer CityELKHART IN 46516
Manufacturer CountryUS
Manufacturer Postal Code46516
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMECHANICAL CHAIR/TRANSPORT CHAIR
Generic NameMEDICAL CHAIR AND TABLE
Product CodeKMN
Date Received2018-07-26
Model NumberUNKNOWN
Catalog NumberUNKNOWN
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCHAMPION MANUFACTURING INC
Manufacturer Address2601 INDUSTRIAL PARKWAY ELKHART IN 46516 US 46516


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-07-26