OPMI SENSERA N/A 000000-1183-793

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-07-27 for OPMI SENSERA N/A 000000-1183-793 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[115404622] A health care professional (hcp) reported that while the doctor was moving the opmi sensera microscope to the operating table prior to a procedure, the optical head became loose from the suspension arm. The head did not fully detach and was still connected to the arm via a cable. The doctor was supporting the head on his hand which prevented it from falling. No patient was in the room at this time. The doctor mounted the optical head back onto the stand using an allen wrench. The same microscope was used to complete the scheduled procedure with minor delay. There was no injury reported to the patient, doctor or a third party.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2018-00010
MDR Report Key7729155
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-07-27
Date of Report2018-06-28
Date of Event2018-06-28
Date Mfgr Received2018-06-28
Device Manufacturer Date2002-11-12
Date Added to Maude2018-07-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPMI SENSERA
Generic NameMICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY
Product CodeFSO
Date Received2018-07-27
Model NumberN/A
Catalog Number000000-1183-793
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.