TRAPEZE FLOOR STAND - DEVICE #1 IC-7741 FLOOR STAND

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2006-10-24 for TRAPEZE FLOOR STAND - DEVICE #1 IC-7741 FLOOR STAND manufactured by Sunrise Medical.

Event Text Entries

[519869] It was reported to mfr, by facility and reporter, per facility they have a 325 lb man using the trapeze for bed mobility / repositioning and not supporting his 325 lb weight. The trapeze triangle was on the floor when they found it, and the resident stated that it hit him in the head, resident did have a bump to his head. Facility also said the state was in and they did a state incident report on this. Per mfr in our user instruction manual, it states under the warnings "the head trapeze, used in conjunction with a bed, can support an individual weighing up to 250 lbs. " facility is stating because we do not have a label on the triangle itself and because we state "can support" and "assists the user when repositioning in bed" facility wants to know what do we mean by that verbage. Facility insists we are not clear, and because the pt was using it for bed mobility / repositioning and not supporting his 325 lbs, they want something in writing from us. Per mfg this is in question; because we believe when a person is repositioning or using a trapeze for bed mobility you are supporting/ utilizing your weight. Per product manager, if the trapeze is properly installed and the triangle locked down, per the mfr installation instructions, there is no way for it to fall off.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182305-2006-00038
MDR Report Key773538
Report Source06
Date Received2006-10-24
Date of Report2006-10-16
Date of Event2006-09-23
Date Facility Aware2006-09-23
Report Date2006-10-16
Date Mfgr Received2006-09-25
Date Added to Maude2006-10-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSHERYL NUGENT
Manufacturer Street5001 JOERNS DR.
Manufacturer CitySTEVENS POINT WI 54481
Manufacturer CountryUS
Manufacturer Postal54481
Manufacturer Phone7153413600
Manufacturer G1SUNRISE MEDICAL
Manufacturer Street5001 JOERNS DR.
Manufacturer CitySTEVENS POINT WI 54481
Manufacturer CountryUS
Manufacturer Postal Code54481
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRAPEZE FLOOR STAND - DEVICE #1
Generic NameTRAPEZE FLOOR STAND BED ACCESSORY
Product CodeILZ
Date Received2006-10-24
Model NumberIC-7741 FLOOR STAND
Catalog NumberIC-7741 FLOOR STAND
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeUNK
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key761322
ManufacturerSUNRISE MEDICAL
Manufacturer Address5001 JOERNS DR. STEVENS POINT WI 54481 US
Baseline Brand NameTRAPEZE FLOOR STAND - DEVICE #1
Baseline Generic NameTRAPEZE FLOOR STAND BED ACCESSORY
Baseline Model NoIC-7741 FLOOR S
Baseline Catalog NoIC-7741 FLOOR STAND
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-10-24