ETEST? VANCOMYCIN VA 256 WW S30 525518

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-07-31 for ETEST? VANCOMYCIN VA 256 WW S30 525518 manufactured by Biomerieux Sa.

Event Text Entries

[115927385] On (b)(6) 2018, a eucast member reported to biom? Rieux a potential publication related to the underestimation of mic value for enterococci with a low level of resistance, in association with etest? Vancomycin va 256 ww s30 (ref. (b)(4)). The eucast development laboratory confirmed the findings from a (b)(6) reference laboratory for resistant enterococci strains (mic value >4 mg/l) with a van b genotype regarding an underestimation of the mics value with etest? Vancomycin va 256 (ref. (b)(4)) that led for some of the isolates to a susceptible result on etest? Instead of resistant. The study compared etest? Vancomycin to the broth micro dilution reference method and the results were: essential agreement: 63%, category agreement: 38%, false susceptible: 63%. No patient outcomes were reported due to the referenced issue. There is no indication or report that the discrepant results led to any adverse event related to any patient's state of health. A biom? Rieux internal investigation has been initiated and impacted strains were requested.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00123
MDR Report Key7735896
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-07-31
Date of Report2018-11-29
Date Mfgr Received2018-10-30
Device Manufacturer Date2017-10-04
Date Added to Maude2018-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318805
Manufacturer G1BIOMERIEUX SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE 383,
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? VANCOMYCIN VA 256 WW S30
Generic NameETEST? VANCOMYCIN VA 256 WW S30
Product CodeJWY
Date Received2018-07-31
Catalog Number525518
Lot Number1006011340
Device Expiration Date2018-10-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE 383, FR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-07-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.