NEUROCOM EQUITEST NCM-SMEQ

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-07-31 for NEUROCOM EQUITEST NCM-SMEQ manufactured by Natus Medical Incorporated.

Event Text Entries

[115650136] Natus technical service confirmed that the system was frozen while troubleshooting the issue remotely with the complainant. Natus technical service walked the complainant through turning off the dynamic platform and disconnecting power to the computer. After letting the system rest for at least a minute, power was reconnected. The system then operated correctly and the failure could no longer be duplicated. Natus technical service advised the complainant to continue monitoring the system and contact natus should the issue reoccur. Because the issue resolved itself after cycling power to the system, the cause of the issue could not be determined.
Patient Sequence No: 1, Text Type: N, H10


[115650137] It was reported to natus that during a gaze stabilization test (gst), the smart equitest system froze. There was no report of death, serious injury, or environmental/safety concerns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00135
MDR Report Key7737403
Report SourceUSER FACILITY
Date Received2018-07-31
Date of Report2018-07-02
Date of Event2018-06-29
Date Mfgr Received2018-07-02
Date Added to Maude2018-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEUROCOM EQUITEST
Generic NameSMART EQUITEST
Product CodeKHX
Date Received2018-07-31
Model NumberNCM-SMEQ
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-07-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.