[115797067] The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[115797068] The customer complained of questionable elecsys vitamin b12 immunoassay results for 11 patient samples tested on a cobas e 411 immunoassay analyzer compared to the results from another site. The specific analyzer used at the other site was not provided, but it was provided that the methodology was electro-chemiluminescence immunoassay (eclia). From the data provided, 5 patient results were a reportable malfunction. Please refer to the attachment to this medwatch for patient data. The erroneous results were released outside of the laboratory. There was no allegation of an adverse event. The cobas e411 serial number was (b)(4). The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-02546
MDR Report Key7739608
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-08-01
Date of Report2018-09-17
Date of Event2018-07-12
Date Mfgr Received2018-07-17
Date Added to Maude2018-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS VITAMIN B12 IMMUNOASSAY
Generic NameRADIOASSAY, VITAMIN B12
Product CodeCDD
Date Received2018-08-01
Model NumberNA
Catalog Number07212771190
Lot Number305578
Device Expiration Date2018-10-31
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-01