OLYMPIC COOL-CAP 401351-505

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-08-01 for OLYMPIC COOL-CAP 401351-505 manufactured by Natus Medical Incorporated.

Event Text Entries

[115776322] The cool-cap system was returned to natus for evaluation. Natus factory service found that a cooling module inside the cooling block was operating out of specification. Both cooling modules were replaced and the repaired cool-cap system was returned to the user facility. The complainant reported that the repaired cool-cap system passed their initial testing. No further communication was received.
Patient Sequence No: 1, Text Type: N, H10


[115776323] It was reported to natus that a cool-cap system was not cooling quickly enough when the cap temperature was set to 8 degrees c. The temperature remained at 14 degrees c three hours after setting the cap temperature to 8 degrees c. A patient was being treated with the system when the event occurred. The patient was immediately moved to another cool-cap system. The cool-cap system involved in the event was taken out of service. There was no report of death, serious injury, or environmental/safety concerns. The biomed at the user facility who reported the event did not witness the event and could not provide any additional details. Temperature data could not be retrieved because the system had been rebooted. The biomed tested the system's performance immediately after the event. The patient weight was selected as (b)(6) and the gestational age was selected as 36 weeks which prompted the cap temperature to 23. 1 degrees c. After an initial cooling period the biomed set the cap temperature to 12 degrees c. The biomed stated that it took about an hour for the cap temperature to reach 12 degrees c. Natus technical service asked the biomed to set the cap temperature to 15 degrees c, and within 8 minutes the cap temperature reached 14. 7 degrees c.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00136
MDR Report Key7740215
Report SourceUSER FACILITY
Date Received2018-08-01
Date of Report2017-01-19
Date of Event2017-01-18
Date Mfgr Received2017-01-19
Date Added to Maude2018-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-08-01
Returned To Mfg2017-02-13
Model Number401351-505
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-01