DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM 9438-06 STT-GF-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2018-08-01 for DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM 9438-06 STT-GF-001 manufactured by Dexcom, Inc..

Event Text Entries

[115789731] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[115789732] Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, a pairing failure between the transmitter, receiver, and smart device occurred. No additional patient or event information is available. No product was provided for evaluation. Data was provided for evaluation. Data review did not confirm the reported event of pairing failure. A root cause could not be determined via data. The reported issue of pairing failure is reportable based on the finding of transmitter failure error.
Patient Sequence No: 1, Text Type: D, B5


[139207993] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[139207994] Product was returned. An exterior physical visual inspection was performed and passed. A voltage measurement was taken and failed at 0 vdc. A share log review was performed and a failure to pair was found. A fatal error was observed for a different tx id (b)(4) and is not related to this investigation. The alleged transmitter has already been scrapped.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004753838-2018-078472
MDR Report Key7741637
Report SourceCONSUMER
Date Received2018-08-01
Date of Report2019-03-04
Date of Event2018-05-11
Date Mfgr Received2019-03-04
Device Manufacturer Date2017-08-04
Date Added to Maude2018-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KIPP DURBIN
Manufacturer Street6340 SEQUENCE DR.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582000200
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Generic NameCONTINUOUS GLUCOSE MONITOR
Product CodePQF
Date Received2018-08-01
Returned To Mfg2018-07-25
Model Number9438-06
Catalog NumberSTT-GF-001
Lot Number6004138
Device Expiration Date2018-03-26
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerDEXCOM, INC.
Manufacturer Address6340 SEQUENCE DR. SAN DIEGO CA 92121 US 92121

Device Sequence Number: 1

Brand NameDEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Generic NamePQF
Product CodePFQ
Date Received2018-08-01
Returned To Mfg2018-07-25
Model Number9438-06
Catalog NumberSTT-GF-001
Lot Number6004138
Device Expiration Date2018-03-26
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDEXCOM, INC.
Manufacturer Address6340 SEQUENCE DR. SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.