OLYMPIC COOL-CAP 401701-501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-08-05 for OLYMPIC COOL-CAP 401701-501 manufactured by Natus Medical Incorporated.

Event Text Entries

[116055506] The engineering manager at the user facility conducted testing of the cool-cap system. In this testing it was determined that the water was correctly reaching set temperature and the screen would properly respond to commands. The most recent annual maintenance was conducted 8 months prior to the event, during which it was reported that the cooling unit's fans were blocked by accumulated dust. Natus technical service asked for additional details regarding the patient, the patient's condition, specifics of the reported temperature changes, and responsiveness of the touchscreen when the problem occurred, but no response was received. Natus technical service advised that they could not recommend any field repairs and the system must be returned to natus for repair. The system was not returned despite multiple requests sent by natus technical service.
Patient Sequence No: 1, Text Type: N, H10


[116055507] It was reported to natus by an engineering manager at the user facility that the temperature of a cool-cap system was changing constantly and no alarm would indicate when these temperature changes would occur. The engineering manager stated that nurses at the user facility claimed that the cool-cap system would restart several times during treatment. The nurses claimed that after one of these restarts the cap could no longer be cooled, the water level in the cap was low, and a patient's temperature rose. There was no report of death, serious injury, or environmental/safety concerns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00137
MDR Report Key7750161
Report SourceUSER FACILITY
Date Received2018-08-05
Date of Report2016-09-14
Date Mfgr Received2016-09-14
Date Added to Maude2018-08-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-08-05
Model Number401701-501
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.