[116049128] Natus technical service made multiple attempts to request additional information regarding the patient and their condition, but these requests were not answered. A technician sent to the user facility verified that the correct mains voltage was reaching the power supply, but the system would not turn on and the power supply fan was not spinning. The user facility was provided with a replacement power supply. It was reported that once the replacement power supply was installed, the system began working correctly and was put back into service.
Patient Sequence No: 1, Text Type: N, H10


[116049129] It was reported to natus by a distributor that a cool-cap system powered down during treatment and treatment was interrupted. Once powered down, the system would no longer turn on. There was no report of death, serious injury, or environmental/safety concerns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00138
MDR Report Key7750206
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-08-05
Date of Report2016-10-27
Date Mfgr Received2016-10-27
Date Added to Maude2018-08-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-08-05
Model Number401701-501
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.