LIPC LIPASE COLORIMETRIC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-06 for LIPC LIPASE COLORIMETRIC manufactured by Roche Diagnostics.

Event Text Entries

[116865174] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[116865175] The customer questioned high results for 1 patient tested for lipase on a cobas 6000 c (501) module. The patient had been tested for lipase on the c501 module initially with a result of 739 u/l. The sample was repeated on "another system" and the result was 733 u/l. On (b)(6) 2018 the patient was tested on the c501 module with a result of 981 u/l. This result was reported outside of the laboratory. The sample was sent to an external laboratory using the abbott method and the result was 150 u/l. The date of the event is either (b)(6) 2018. Clarification on this has been requested but not provided. There was no allegation that an adverse event occurred. The c501 module serial (b)(4). The customer suspects an interference in the sample affecting the results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-02615
MDR Report Key7751778
Date Received2018-08-06
Date of Report2018-11-30
Date of Event2018-07-12
Date Mfgr Received2018-07-19
Date Added to Maude2018-08-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameLIPC LIPASE COLORIMETRIC
Generic NameLIPASE TEST SYSTEM
Product CodeCHI
Date Received2018-08-06
Model NumberNA
Catalog NumberASKU
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.