HUDSON VOLDYNE2500 VOLUMETRIC EXERCISER 8884719025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2018-08-06 for HUDSON VOLDYNE2500 VOLUMETRIC EXERCISER 8884719025 manufactured by Teleflex Medical.

Event Text Entries

[116214381] (b)(4). The device involved in this complaint has not been received by the manufacturer for evaluation at the time of this report. Verification of failure mode reported in the current manufacturing process was conducted. Thirteen device samples were taken from the current production. The samples were functionally inspected, and during testing, the issue reported was not observed in the current manufacturing process. A device history record investigation did not show issues related to this complaint. Customer complaint cannot be confirmed due the product sample is not available to perform a proper investigation and determine the root cause. Corrective actions cannot be established at this time. If the device sample becomes available at a later date, this report will be updated accordingly.
Patient Sequence No: 1, Text Type: N, H10


[116214382] Customer complaint alleges the yellow indicator peg sticks during use. Reported by a home care user. Customer states the device must be shaken to reset the peg. No harm was reported. Patient condition reported as "fine".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003898360-2018-00622
MDR Report Key7753472
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2018-08-06
Date of Report2018-07-25
Date of Event2017-12-01
Date Mfgr Received2018-07-25
Device Manufacturer Date2013-10-09
Date Added to Maude2018-08-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA WOODALL
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9196942566
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetRANCHO EL DESCANSO
Manufacturer CityTECATE 21478
Manufacturer CountryMX
Manufacturer Postal Code21478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON VOLDYNE2500 VOLUMETRIC EXERCISER
Generic NameSPIROMETER, THERAPEUTIC (INCEN
Product CodeBWF
Date Received2018-08-06
Catalog Number8884719025
Lot Number01J1300525
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-06