BIB CATHETER 420.1 BB072

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-07 for BIB CATHETER 420.1 BB072 manufactured by Numed, Inc..

Event Text Entries

[116714098] Numed, inc. Was made aware of incident from a returned post market clinical follow-up form. Incident was not reported by hospital. Physician did not think the rupture necessitated a formal complaint. Catheter was not returned. Balloon burst can not be confirmed. A comparative catheter from a different lot was pulled and tested for rated burst pressure. The labeled rbp for this device is 2. 0 atm. The comparative catheters outer balloon was inflated in 0. 5 atm increments until rbp (2. 0 atm) with no issues. It was then over inflated to 5. 0 atm, which is more than double the rbp, where it burst longitudinally. An inflation device with pressure gauge was used. Procedure was successful. Very little information has been received, even after repeated attempts.
Patient Sequence No: 1, Text Type: N, H10


[116714099] Rupture of the outer balloon at 2 atm and full diameter. Withdrawal without problems.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1318694-2018-00010
MDR Report Key7757950
Date Received2018-08-07
Date of Report2018-08-07
Date of Event2017-11-23
Date Mfgr Received2018-07-13
Device Manufacturer Date2016-12-22
Date Added to Maude2018-08-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MELISSA THOMAS
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal12965
Manufacturer Phone3153284491
Manufacturer G1NUMED, INC.
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal Code12965
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBIB CATHETER
Generic NameSTENT PLACEMENT CATHETER
Product CodeNVM
Date Received2018-08-07
Model Number420.1
Catalog NumberBB072
Lot NumberBIB-8895
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNUMED, INC.
Manufacturer Address2880 MAIN STREET HOPKINTON NY 12965 US 12965


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.