BACT/ALERT? BPN CULTURE BOTTLES 279045

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-08-07 for BACT/ALERT? BPN CULTURE BOTTLES 279045 manufactured by Biomerieux Inc..

Event Text Entries

[116909081] A customer in (b)(6) notified biom? Rieux of poor clostridium sporogenes growth in association with the bact/alert? Bpn culture bottles (reference 279045). The customer reported that as part of each batch of biom? Rieux media bottles they preform growth promotion studies to demonstrate the bottles are capable of growing an array of microbes. The customer stated that they successfully grew all microbes in the referenced lot with the exception of clostridium sporogenes. This returned no growth in all three (3) bottles inoculated. The customer was uncertain if the poor growth was due to poor inoculation, so they repeat tested on another five bottles and received the same result of no growth. These bottles were not used for patient samples, these bottles were for quality control checks only. Since the quality control checks failed the lot was not used for routine testing. Biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00134
MDR Report Key7759727
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-08-07
Date of Report2018-08-07
Date Mfgr Received2018-04-24
Device Manufacturer Date2018-08-25
Date Added to Maude2018-08-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIFFANY HALL
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX INC.
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal Code27712
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT? BPN CULTURE BOTTLES
Generic NameBACT/ALERT? BPN CULTURE BOTTLES
Product CodeMZC
Date Received2018-08-07
Catalog Number279045
Lot Number1049460
Device Expiration Date2018-08-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX INC.
Manufacturer Address100 RODOLPHE STREET DURHAM NC 27712 US 27712


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-07