MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-08-08 for ENDOVIVE? STANDARD PEG KIT M00509081 manufactured by Boston Scientific - Spencer.
[116386682]
(b)(4). The device has been received for analysis. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10
[116386683]
It was reported to boston scientific corporation that an endovive standard peg kit pull method was used during a percutaneous endoscopic gastrostomy (peg) procedure performed on (b)(6) 2018. According to the complainant, during the procedure, as the peg tube device was put in place, the peg tube came out along with its internal bolster. A nasogastric tube was placed to complete the procedure. There were no patient complications reported as a result of this event. The patient's condition was reported to be stable.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005099803-2018-02513 |
MDR Report Key | 7761448 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2018-08-08 |
Date of Report | 2018-07-13 |
Date of Event | 2018-07-10 |
Date Mfgr Received | 2018-08-20 |
Device Manufacturer Date | 2018-04-18 |
Date Added to Maude | 2018-08-08 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | EMP. NANCY CUTINO |
Manufacturer Street | 100 BOSTON SCIENTIFIC WAY |
Manufacturer City | MARLBOROUGH MA 01752 |
Manufacturer Country | US |
Manufacturer Postal | 01752 |
Manufacturer Phone | 5086834000 |
Manufacturer G1 | BOSTON SCIENTIFIC - SPENCER |
Manufacturer Street | 780 BROOKSIDE DRIVE |
Manufacturer City | SPENCER IN 47460 |
Manufacturer Country | US |
Manufacturer Postal Code | 47460 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ENDOVIVE? STANDARD PEG KIT |
Generic Name | GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS |
Product Code | PIF |
Date Received | 2018-08-08 |
Returned To Mfg | 2018-08-01 |
Model Number | M00509081 |
Lot Number | 22007923 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BOSTON SCIENTIFIC - SPENCER |
Manufacturer Address | 780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-08-08 |