MAUDE MDR 7764842

MDR report key
7764842
Report number
1319808-2018-00022
Event key
0
Event type
3
Date of event
2018-07-16
Date received
2018-08-08
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MR. JAMES A STEVENS
Address
100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US
Phone
585-585-5854
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 2IN-VITRO DIAGNOSTICORTHO-CLINICAL DIAGNOSTICSJIX16626590267Y Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12018-08-080

Event Narratives#

N

Patient 1

THE INVESTIGATION DETERMINED THAT A LOWER THAN EXPECTED VITROS NA+ RESULT WAS OBTAINED WHEN PERFORMING A PATIENT SAMPLE CORRELATION ON A VITROS 4600 CHEMISTRY SYSTEM. THE LIKELY CAUSE OF LOWER THAN EXPECTED VITROS NA+ RESULT IS A SUBOPTIMAL CALIBRATION. HOWEVER, THE CAUSE OF THE SUBOPTIMAL CALIBRATION COULD NOT BE DEFINITIVELY DETERMINED. A DIAGNOSTIC WITHIN RUN VITROS NA+ PRECISION TEST WAS NOT PERFORMED TO EVALUATE THE PERFORMANCE OF THE VITROS 4600 SYSTEM, THEREFORE, THE VITROS 4600 SYSTEM COULD NOT BE ENTIRELY RULED OUT AS A CONTRIBUTOR OF THE EVENT. AS VITROS NA+ LOT 4219-0991-0597 HAD NOT BEEN PUT INTO USE, HISTORICAL VITROS NA+ QUALITY CONTROL RESULTS WERE NOT AVAILABLE TO EVALUATE THE PERFORMANCE OF VITROS NA+ LOT 4219-0991-0597. THEREFORE AN ISSUE WITH THE REAGENT COULD NOT BE ENTIRELY RULED OUT AS A CONTRIBUTOR OF THE EVENT. THE LOWER THAN EXPECTED VITROS NA+ RESULT WAS LIKELY CAUSED BY A SUBOPTIMAL CALIBRATION EVENT ON (B)(6) 2018 LIKELY DUE TO A CALIBRATOR HANDLING ISSUE. ACCEPTABLE PERFORMANCE WAS OBSERVED USING FRESHLY PREPARED CALIBRATORS WITHOUT ALTERING THE CONDITION OF THE VITROS 4600 SYSTEM. THIS WOULD INDICATE THAT BOTH VITROS NA+ LOT 4219-0991-0597 AND THE VITROS 4600 SYSTEM DID NOT CONTRIBUTE TO THE SUBOPTIMAL CALIBRATION.

D

Patient 1

A CUSTOMER OBTAINED A LOWER THAN EXPECTED VITROS SODIUM (NA+) RESULT WHEN PERFORMING A PATIENT SAMPLE CORRELATION ON A VITROS 4600 CHEMISTRY SYSTEM. PATIENT SAMPLE 20 VITROS NA+ RESULT 144 MMOL/L VERSUS THE EXPECTED VITROS NA+ RESULT 151 MMOL/L. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THE EVENT WERE TO OCCUR UNDETECTED. THE LOWER THAN EXPECTED NA+ RESULT WAS OBTAINED WHILE PERFORMING A PATIENT SAMPLE CORRELATION AND WAS NOT REPORTED OUTSIDE OF THE LABORATORY. HOWEVER THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. (B)(4).