CARE-E-VAC 3 758000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-08 for CARE-E-VAC 3 758000 manufactured by Ohio Medical Llc.

Event Text Entries

[117003167] Field correction 1419185-7/24/18-001-c has been initiated for correction of other potentially affected devices.
Patient Sequence No: 1, Text Type: N, H10


[117003168] On 11/15/2017 a complaint was received for the care-e-vac 3 device in which the customer stated "out of box failure". Upon return to the manufacturer on 12/7/2017 the following was noted by the repair technician " note: this is not an out of box failure. Evidence of extensive use and seal broken indicating unit has been opened. Inspection in repair center shows this unit had liquid damage causing catastrophic electrical failure. " pictures were provided that show melted wires from the device battery to the pc board. Device was retained by engineering. On (b)(6) 2018 there was an occurence in house during assembly of the care-e-vac 3 device in which same time of effect occured (melted wires leading from battery to the pc board). Ship hold was initiated immediately and investigation began looking further into the cause of the burning/ melted wires. This device was re-evaluated by engineering with comparison to failure which occured in house. Same effects of failure were observed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419185-2018-00002
MDR Report Key7765710
Date Received2018-08-08
Date of Report2018-08-08
Date of Event2018-06-22
Device Manufacturer Date2017-06-16
Date Added to Maude2018-08-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA BARRILE
Manufacturer Street1111 LAKESIDE DR
Manufacturer CityGURNEE IL 60031
Manufacturer CountryUS
Manufacturer Postal60031
Manufacturer Phone8478556318
Manufacturer G1OHIO MEDICAL LLC
Manufacturer Street1111 LAKESIDE DR
Manufacturer CityGURNEE IL 60031
Manufacturer CountryUS
Manufacturer Postal Code60031
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCARE-E-VAC 3
Generic NamePORTABLE SUCTION DEVICE
Product CodeBTA
Date Received2018-08-08
Returned To Mfg2017-12-07
Model Number758000
Catalog Number758000
Lot NumberCEB171029
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOHIO MEDICAL LLC
Manufacturer Address1111 LAKESIDE DR GURNEE IL 600312489 US 600312489


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.